Requirements for optimal placement conditions of ventilating aging test chamber

Ventilation aging test chamber can work normally under the following conditions:

1. Temperature: 5 ~ 40 ℃;

2. Relative humidity: not more than 85%;

3. Air pressure: 86 ~ 106kPa;

4. Power supply: AC voltage 220 ± 22V or 380 ± 38V, frequency 50 ± 0.5Hz;

5. There is no strong vibration around, no strong magnetic field, no corrosive gas;

6. No direct sunlight or direct radiation from other cold and heat sources;

8. There is no strong airflow around. When the surrounding air needs forced convection, the airflow should not blow directly onto the box.

Note: The above is the best placement condition of the ventilating aging test box. If the conditions are not allowed, please go to the Beijing Yashilin Technology Department. Beijing Yashilin will serve you wholeheartedly.